How do I reconcile 96.4% from Janoshik with 98.8% from the supplier?

Asked 16 Oct 2025Modified 6 months agoViewed 12k times
23

What I have: 96.4% · Janoshik · 98.8%.

Before I write this off, I want to check whether it is a known failure mode.

I have the lot number, the certificate and the date, and I am happy to compare them against anything.

What is the differential here, and which test discriminates between the options?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
lc-ms
lc-ms

Liquid chromatography coupled to mass spectrometry: separation and identification in one run, and the method of choice when you need to know not…

14 questions
shareeditfollowflag
KN
askedklara_novotna16k1616 Oct 2025
5Good answer, but the confidence interval in the cited trial is wider than implied. – Dr_Colm_Fitzhenry 2 months ago
add a comment

5 Answers

Accepted answer first, then by votes
57

Accepted answer

Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

Reconciling gross mass to label claim

ComponentTypical shareCounted in purity?Counted in content?
Target peptide88–94 %Yes, as main peakYes
Related impurities1–3 %Yes, as other peaksNo
Counter-ion (TFA or acetate)2–8 %NoNo
Residual water2–6 %NoNo
Bulking agent, if present0–40 %NoNo

To be exact about it, column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
DV
answered · accepteddead_volume49k3810 Jan 2026
4Good answer, but the confidence interval in the cited trial is wider than implied. – triple_agonist_q 3 months ago
add a comment
Sponsored

Janoshik Analytical - Independent Third-Party Testing

HPLC purity, identity confirmation and quantified content on the vial you actually hold. Reports arrive with the chromatogram attached, not just a number.

Submit a sample
Sponsored — paired listing

GL Biochem (Shanghai) Ltd. - Direct Synthesis

Founded 1998. ISO 9001 and cGMP certified, 1,500+ staff and 200+ patents. The synthesis house behind a great many of the vials that get sent out for testing - batch-specific documentation with every order.

Visit GL Biochem
63

To be exact about it, gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.

Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

Compare purity within a single laboratory on the same method, never across laboratories.

edited 4 Jan 2026 by h_villanueva — removed a claim I could not source

shareimprove this answerflag
HV
answeredh_villanueva50k3819 Dec 2025
43

It helps to be literal here: read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

The relevant detail is that buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
M4
answeredmz_411399k2588 Dec 2025
27

The short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
CL
answeredcold_lane14k1830 Dec 2025
20

The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.

Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag