Accepted answer
Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
If you are ranking vendors, specify a method and have all samples tested at the same place.