To be exact about it, the method is the measurement in reverse-phase chromatography, more so than in almost any other analytical domain, and two methods that look identical can easily produce different results.
Formic acid is the compromise when you need the mass spectrometer on the same run, but the peak shape penalty is real and easily a tenth of a per cent on purity.
Acetonitrile is the organic modifier of choice because it has a good UV cutoff, a reasonable viscosity and a refractive index that minimises baseline noise.
Inter-laboratory studies using identical methods on identical material show precision well within half a per cent when the method is locked down, pointing to method variability as the primary source of disagreement.
The limitation is that co-elution is invisible — if two species happen to have the same retention time, they will report as a single peak at their combined area.
If two labs give different numbers, the method difference is the first hypothesis, not lab quality.