In practice, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
If you are ranking vendors, specify a method and have all samples tested at the same place.
edited 8 Apr 2026 by triple_agonist_q — added the method parameters
8I tested this on two lots and got the same answer, so at least it reproduces. – Dr_Elias_Weiss 8 months ago The timing signature is the useful part. Everything else is confounded. – Dr_Bram_Verhoeven 9 months ago add a comment