System suitability is the part of a report that tells you whether the method was under control on the day you were tested, and its absence is concerning.
Sample preparation is almost always under-appreciated — a reconstituted peptide in strong solvent will distort its own peak on the gradient.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
Put another way, wider-pore phases — 300 Angstrom rather than 100 Angstrom — have faster mass transfer and narrower peaks for peptides above three kilodaltons, which is almost every peptide you will see.
The resolving power of a separation is quantified by the resolution parameter R, defined from the heights and widths of adjacent peaks, and pharmacopoeial methods typically demand R greater than 1.5 for a method to be considered validated.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
Ask for the chromatogram and the system suitability data, not just the number.