The underlying point is that purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.