99.4 per cent is a statement about area, and the other 0.6 per cent is everything the detector saw and did not assign to your peak. Read it as 99.4 of every 100 units of peak area at whatever wavelength was used, not as 99.4 per cent of the mass in the vial. With the batch certificate attached you can at least see how the figure was produced, which is the difference between a measurement and a claim. What it still does not tell you is content: how many milligrams are actually there.
Start from what the detector sees, because that tells you what the number means.
Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
If you are ranking vendors, specify a method and have all samples tested at the same place.