Accepted answer
It helps to be literal here: degradation is not one process, and which one dominates depends on the condition you are asking about. In solution at refrigerated temperature the rate-limiting pathway is usually deamidation and hydrolysis; at room temperature aggregation overtakes them; frozen, the damage happens during the transitions rather than during the hold.
Light matters for specific residues rather than in general. Tryptophan and to a lesser extent tyrosine and methionine are photo-labile; a sequence without them is largely indifferent to ambient light over the timescales in question. Amber glass is cheap insurance rather than a requirement.
Degradation pathway by condition
| Pathway | Dominant when | Detected by |
|---|
| Deamidation | Solution, neutral to alkaline pH | RP-HPLC, +1 Da on MS |
| Oxidation | Light, trace metals, peroxides | RP-HPLC, +16 Da on MS |
| Hydrolysis | Solution, extremes of pH | RP-HPLC, fragment masses |
| Aggregation | Agitation, interfaces, high concentration | SEC, visual haze; often invisible on RP-HPLC |
| Freeze-concentration damage | Freeze-thaw of buffered solution | SEC, loss of recovered content |
In practice, aggregation is the failure mode that reverse-phase HPLC is worst at detecting, because a large soluble aggregate may not elute at all and an insoluble one is filtered out during sample preparation. If your purity result comes back normal but the vial looks hazy, believe the vial. Size-exclusion chromatography is the method that sees this.
Deamidation kinetics for asparagine in peptides are well characterised and strongly sequence-dependent: the residue following the asparagine dominates the rate, with glycine and serine at the n+1 position accelerating it by an order of magnitude relative to bulkier residues. That is why two peptides in the same buffer at the same temperature can have quite different shelf lives.
I would be careful about generalising across sequences. Stability is sequence-specific, and a rule derived from semaglutide will not transfer cleanly to a tri-agonist with different residues in different local environments.
If the material arrived warm and it was lyophilised, test it and proceed on the result. If it arrived warm and it was in solution, the result is more likely to be interesting than reassuring.