In practice, reverse-phase HPLC is the workhorse for peptide purity work, but it is almost universally run under conditions that are not optimal for a peptide of this chain length.
Wider-pore phases — 300 Angstrom rather than 100 Angstrom — have faster mass transfer and narrower peaks for peptides above three kilodaltons, which is almost every peptide you will see.
Trifluoroacetic acid at 0.1 per cent is the standard ion-pairing agent because it suppresses tailing by neutralising the basic residues, but it also suppresses mass spectrometry signal.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
If two labs give different numbers, the method difference is the first hypothesis, not lab quality.