Gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
In practice, buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
If you are ranking vendors, specify a method and have all samples tested at the same place.