The single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
Compare purity within a single laboratory on the same method, never across laboratories.