The most important parameter is the one that is almost never specified: the gradient slope during the region where your main peak elutes.
Trifluoroacetic acid at 0.1 per cent is the standard ion-pairing agent because it suppresses tailing by neutralising the basic residues, but it also suppresses mass spectrometry signal.
Stated carefully, detector linearity is why a heavily loaded peak can give false purity by compressing the main peak height while leaving impurity shoulders unchanged.
Peptide impurities from solid-phase synthesis include deletion sequences, truncations from premature cleavage, racemised residues from epimerisation and oxidised variants, each of which may have different chromatographic behaviour.
The caveat is that HPLC is a purity technique and says almost nothing about whether the main peak is actually your target compound — that is why identity confirmation from mass spectrometry or peptide mapping matters.
If two labs give different numbers, the method difference is the first hypothesis, not lab quality.