Degradation is not one process, and which one dominates depends on the condition you are asking about. In solution at refrigerated temperature the rate-limiting pathway is usually deamidation and hydrolysis; at room temperature aggregation overtakes them; frozen, the damage happens during the transitions rather than during the hold.
Aggregation is the failure mode that reverse-phase HPLC is worst at detecting, because a large soluble aggregate may not elute at all and an insoluble one is filtered out during sample preparation. If your purity result comes back normal but the vial looks hazy, believe the vial. Size-exclusion chromatography is the method that sees this.
To be exact about it, for the solid state, residual moisture is the dominant variable. A cake at two per cent water is considerably more stable than the same cake at six per cent, because water is both a reactant in hydrolysis and a plasticiser that lowers the glass transition temperature. This is why a desiccant in the outer packaging is not theatre, and why opening a cold vial in a humid room is a genuine error — you condense water onto the cake.
The caveat is that "within specification" and "unchanged" are different claims. A vial can lose a few per cent of content and still be usable for its purpose while no longer matching its certificate.
Store solid, store cold, store dry, and reconstitute what you will use rather than what fits in the vial.
Good answer, but the confidence interval in the cited trial is wider than implied. – tobias_maartens 6 months ago add a comment