Mechanically, high-performance liquid chromatography is a separation technique, not an identification technique, and it answers the question "how much of the detected signal is a single species" not "which species is it."
Sample preparation is almost always under-appreciated — a reconstituted peptide in strong solvent will distort its own peak on the gradient.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
Reverse-phase stationary phases use C18 or C8 chains bonded to silica, and the pore size of the silica matters more for a peptide of this chain length than the ligand length does.
Peptide separation by reverse-phase high-performance liquid chromatography is described in the general chapters of the United States Pharmacopeia, European Pharmacopeia and Japanese Pharmacopeia, and the methods converge on essentially the same principles.
The practical summary: specify the method, run the same method on every sample you compare, and use orthogonal techniques to confirm the result.