Gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.
Compare purity within a single laboratory on the same method, never across laboratories.