99.1 per cent is at the top of what stepwise synthesis delivers on a chain this long, and it is reachable rather than fictional. Every coupling is high-yielding and none is quantitative, so the deletion and truncation sequences that survive purification are what occupies the remaining 0.9 per cent. Above roughly 98 per cent you are fighting the purification rather than the synthesis, which is why a 99.1 per cent figure on tirzepatide deserves a method question — column, gradient, wavelength — rather than either belief or dismissal.
The single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
Worth being precise here: tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
Compare purity within a single laboratory on the same method, never across laboratories.
2Any reason to prefer ion chromatography over fluorine NMR for the counter-ion here? – sasha_ferreira 6 months ago Worth adding that the method section is where the answer usually is. – tobias_maartens 4 months ago add a comment