Accepted answer
97.9 per cent is at the top of what stepwise synthesis delivers on a chain this long, and it is reachable rather than fictional. Every coupling is high-yielding and none is quantitative, so the deletion and truncation sequences that survive purification are what occupies the remaining 2.1 per cent. Above roughly 98 per cent you are fighting the purification rather than the synthesis, which is why a 97.9 per cent figure on survodutide deserves a method question — column, gradient, wavelength — rather than either belief or dismissal.
Read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
To be exact about it, detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.