The single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
It helps to be literal here: mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
If you are ranking vendors, specify a method and have all samples tested at the same place.
edited 28 Jun 2026 by Dr_Nadia_Farsi — tightened the wording; no substantive change
Related: the same reasoning applies to the counter-ion question. – Dr_Yusuf_Adeyemi 8 months ago add a comment