Concretely, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."
If you are ranking vendors, specify a method and have all samples tested at the same place.