98.8 per cent is a statement about area, and the other 1.2 per cent is everything the detector saw and did not assign to your peak. Read it as 98.8 of every 100 units of peak area at whatever wavelength was used, not as 98.8 per cent of the mass in the vial. With the batch certificate attached you can at least see how the figure was produced, which is the difference between a measurement and a claim. What it still does not tell you is content: how many milligrams are actually there.
Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
Compare purity within a single laboratory on the same method, never across laboratories.