Accepted answer
On the detail: the single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.
Compare purity within a single laboratory on the same method, never across laboratories.