Accepted answer
The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
In practice, buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
edited 25 Dec 2024 by rhian_prydderch — clarified the distinction between purity and content