The underlying point is that high-performance liquid chromatography is a separation technique, not an identification technique, and it answers the question "how much of the detected signal is a single species" not "which species is it."
System suitability checks on replicate injections of a known standard establish whether the method was in control — if the peak area varies by more than two per cent between replicates, something is wrong.
Acetonitrile is the organic modifier of choice because it has a good UV cutoff, a reasonable viscosity and a refractive index that minimises baseline noise.
The selectivity of a reverse-phase separation depends on the hydrophobicity of the side-chain profile, which is why two peptides can co-elute even if they differ by a residue.
The limitation is that co-elution is invisible — if two species happen to have the same retention time, they will report as a single peak at their combined area.
Ask for the chromatogram and the system suitability data, not just the number.