The part that matters: the short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Worth being precise here: tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
Compare purity within a single laboratory on the same method, never across laboratories.