The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
If you are ranking vendors, specify a method and have all samples tested at the same place.