How do I reconcile 99.1% from Janoshik with 96.4% from the supplier?

Asked 27 Apr 2025Modified 11 months agoViewed 19k times
15

What I have: 99.1% · Janoshik · 96.4%.

I have a result I cannot explain, and I would rather diagnose it than guess.

I have checked the obvious explanations and eliminated the two easiest ones.

What is the differential here, and which test discriminates between the options?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
shareeditfollowflag
DC
askedDr_Idris_Coulibaly40k13827 Apr 2025

3 Answers

Sorted by votes
82

Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.

Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.

Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
TI
answeredteodora_ilic15k2828 Apr 2025
Useful. I have added the accept threshold suggestion to my own notes. – tobias_maartens 2 months ago
add a comment
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards
55

The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

Specifically, mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

If you are ranking vendors, specify a method and have all samples tested at the same place.

edited 11 Sept 2025 by Dr_Nadia_Farsi — added the citation requested in comments

shareimprove this answerflag
DF
answeredDr_Nadia_Farsi90k25815 Aug 2025
43

Start from what the detector sees, because that tells you what the number means.

Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

The part that matters: the fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
LS
answeredlow_dead_space42k3820 May 2025