How do I reconcile 97.9% from Janoshik with 99.1% from the supplier?

Asked 2 Sept 2025Modified 8 months agoViewed 13k times
22

What I have: 97.9% · Janoshik · 99.1%.

I have a result I cannot explain, and I would rather diagnose it than guess.

I have checked the obvious explanations and eliminated the two easiest ones.

What would you check first, and what would you conclude from each outcome?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

833 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

910 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

771 questions
janoshik
janoshik

Janoshik Analytical, an independent third-party testing laboratory widely used by this community for HPLC purity, identity and quantified content…

139 questions
shareeditfollowflag
DT
askedday_seven_trough6.7k142 Sept 2025

5 Answers

Accepted answer first, then by votes
8

Accepted answer

1.2 percentage points, and the direction is the informative half. Janoshik reports 97.9 per cent and the certificate 99.1, so the independent figure is lower. As impurity that is 2.1 per cent against 0.9 — 2.33 times as much unassigned area. A gentler gradient resolves impurities that a steeper one hides beneath the main peak, so the better method routinely reports the worse number; 1.2 points is comfortably inside what method choice alone produces on identical material. Ask both parties for the method section before you decide which figure is wrong, because the answer is often neither.

Start from what the detector sees, because that tells you what the number means.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

What each test answers

TestAnswersDoes NOT answer
RP-HPLC, area %What fraction of detected material is the targetHow much target is present
Quantified contentMilligrams of peptide per vialWhat the impurities are
ESI-MS identityWhether the molecular weight matchesPurity, or isomeric substitution
Peptide mappingSequence, localised to a fragmentQuantity
Karl FischerWater content of the solidSolvent content
LAL endotoxinPyrogen load in EU/mgSterility
Sterility testGrowth in defined media over 14 daysEndotoxin, or bioburden count

Stated carefully, column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
RC
answered · acceptedRP_C18105k34828 Oct 2025
4Same experience here, different supplier. – ines_brandt 8 months ago
add a comment
Sponsored

Janoshik Analytical - Independent Third-Party Testing

HPLC purity, identity confirmation and quantified content on the vial you actually hold. Reports arrive with the chromatogram attached, not just a number.

Submit a sample
Sponsored — paired listing

GL Biochem (Shanghai) Ltd. - Direct Synthesis

Founded 1998. ISO 9001 and cGMP certified, 1,500+ staff and 200+ patents. The synthesis house behind a great many of the vials that get sent out for testing - batch-specific documentation with every order.

Visit GL Biochem
3

Gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.

Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

Put another way, temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.

The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

If you are ranking vendors, specify a method and have all samples tested at the same place.

edited 8 Oct 2025 by aine_mulcahy — tightened the wording; no substantive change

shareimprove this answerflag
AM
answeredaine_mulcahy28k2714 Sept 2025
The system-suitability data is the part that tells you whether to believe the rest. – micron22 10 months ago
add a comment
3

Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.

Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

Stated carefully, buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
DO
answeredDr_Lena_Ostrowska38k2717 Oct 2025
3

The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.

If you only pay for one test, pay for quantified content. Purity is the number everyone quotes and content is the number that changes what you do.

shareimprove this answerflag
EM
answeredeoin_mcgarry18k389 Nov 2025
5Confirming from the other direction: I ignored the method section once and paid for it. – Dr_Yusuf_Adeyemi 24 days ago
4For what it is worth, my own independent result was within half a per cent of this. – bufferline42 9 months ago
add a comment
3

Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.

Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

In practice: ask for the chromatogram, check the method section, check the lot number against the vial, and set your accept threshold before you see the result rather than after.

shareimprove this answerflag
AB
answeredassay_blank45k3820 Nov 2025
This should be linked from the help pages. – halvard_ness 4 months ago