High-performance liquid chromatography is a separation technique, not an identification technique, and it answers the question "how much of the detected signal is a single species" not "which species is it."
Proline-rich sequences are particularly problematic because the isomerisation kinetics are in the same timescale as the separation, leading to split or broadened peaks at low temperature.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
The relevant detail is that mobile phase pH at the point where you inject must match the mobile phase pH at the start of the gradient, or the sample will not be focused at the column head.
Peptide separation by reverse-phase high-performance liquid chromatography is described in the general chapters of the United States Pharmacopeia, European Pharmacopeia and Japanese Pharmacopeia, and the methods converge on essentially the same principles.
The practical summary: specify the method, run the same method on every sample you compare, and use orthogonal techniques to confirm the result.