Is a 99.4% result from Janoshik comparable to 97.1% from Medutest?

Asked 22 Feb 2025Modified 14 months agoViewed 10k times
11

Details up front: 99.4% · Janoshik · 97.1% · Medutest.

I want to know what the trade-off actually is rather than which option is fashionable.

I would rather have a defensible reason than a marginal improvement.

So which one, and on what grounds?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

833 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

910 questions
reference-standard
reference-standard

Certified reference materials and in-house standards: traceability, the certificate of the standard itself, system suitability, and why every…

185 questions
janoshik
janoshik

Janoshik Analytical, an independent third-party testing laboratory widely used by this community for HPLC purity, identity and quantified content…

139 questions
medutest
medutest

Medutest, one of four independent testing and verification services referenced throughout this site. Use this tag for its methods, its report…

126 questions
shareeditfollowflag
MI
askedmicron2222k3822 Feb 2025

5 Answers

Accepted answer first, then by votes
20

Accepted answer

99.4 and 97.1 are 2.3 percentage points apart, which sounds small until you restate it as impurity. Janoshik is leaving 0.6 per cent of the detected area unassigned and Medutest 2.9 per cent — a factor of 4.83 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 99.4 and 97.1 are two measurements of slightly different quantities that happen to share a unit.

Start from what the detector sees, because that tells you what the number means.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

What each test answers

TestAnswersDoes NOT answer
RP-HPLC, area %What fraction of detected material is the targetHow much target is present
Quantified contentMilligrams of peptide per vialWhat the impurities are
ESI-MS identityWhether the molecular weight matchesPurity, or isomeric substitution
Peptide mappingSequence, localised to a fragmentQuantity
Karl FischerWater content of the solidSolvent content
LAL endotoxinPyrogen load in EU/mgSterility
Sterility testGrowth in defined media over 14 daysEndotoxin, or bioburden count

Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
AL
answered · accepteda_lindgren58k24820 May 2025
8Do you have the chromatogram for this, or just the summary figure? – rae_oyelowo 7 months ago
add a comment
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards
17

Gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

Mechanically, sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
AL
answereda_lindgren58k24828 Apr 2025
3Thank you — this is the answer I was looking for. – j_wierzbicki 6 months ago
add a comment
13

Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.

Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

More usefully, mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
JV
answeredjo_vandeberg23k2817 Apr 2025
9

Reporting threshold is convention and not chemistry, which is why two certificates with different thresholds disagree by a tenth of a point or more.

Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

If you only pay for one test, pay for quantified content. Purity is the number everyone quotes and content is the number that changes what you do.

shareimprove this answerflag
NN
answerednine_point_nine60k1489 May 2025
Same experience here, different supplier. – tobias_reint 10 months ago
Which wavelength was the purity integrated at? It changes the number more than people think. – Dr_Aoife_Brennan 43 days ago
add a comment
8

The honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.

In practice: ask for the chromatogram, check the method section, check the lot number against the vial, and set your accept threshold before you see the result rather than after.

edited 30 Mar 2025 by sample_id — removed a claim I could not source

shareimprove this answerflag
SI
answeredsample_id17k2714 Mar 2025
7The system-suitability data is the part that tells you whether to believe the rest. – mz_4113 2 months ago
add a comment