Coupling the chromatography to mass spectrometry adds identity information but trades chromatographic quality for ionisation efficiency.
Trifluoroacetic acid at 0.1 per cent is the standard ion-pairing agent because it suppresses tailing by neutralising the basic residues, but it also suppresses mass spectrometry signal.
Detector linearity is why a heavily loaded peak can give false purity by compressing the main peak height while leaving impurity shoulders unchanged.
The selectivity of a reverse-phase separation depends on the hydrophobicity of the side-chain profile, which is why two peptides can co-elute even if they differ by a residue.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
If two labs give different numbers, the method difference is the first hypothesis, not lab quality.