Accepted answer
Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
In practice, temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
I would gently push back on the second point — the evidence there is thinner than stated. – dead_volume 18 days ago 8Adding for future readers: the certificate should carry the lot number, not just a batch code. – swirl_dont_shake 9 months ago add a comment