Accepted answer
In practice, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Worth being precise here: mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
Compare purity within a single laboratory on the same method, never across laboratories.
Two of us worked through this independently and arrived here, so it is at least reproducible. – zeynep_arslan 6 months ago Worth adding that the method section is where the answer usually is. – Dr_Nadia_Farsi 4 months ago add a comment