Concretely, gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.