How do I reconcile 97.9% from Medutest with 98.4% from the supplier?

Asked 12 Jul 2026Modified 17 hours agoViewed 3.7k times
14

What I am working with: 97.9% · Medutest · 98.4%.

I noticed this today and I have not touched anything since, in case the state is diagnostic.

I have not discarded anything yet, so a test is still possible if that is the recommendation.

Should I be treating this as a failure or as noise?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
shareeditfollowflag
MF
askedmeniscus_film34k3812 Jul 2026

5 Answers

Sorted by votes
36

Stated carefully, start from what the detector sees, because that tells you what the number means.

Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

Reconciling gross mass to label claim

ComponentTypical shareCounted in purity?Counted in content?
Target peptide88–94 %Yes, as main peakYes
Related impurities1–3 %Yes, as other peaksNo
Counter-ion (TFA or acetate)2–8 %NoNo
Residual water2–6 %NoNo
Bulking agent, if present0–40 %NoNo

Mechanically, integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
P9
answeredplate_count_9k95k15819 Jul 2026
5I would gently push back on the second point — the evidence there is thinner than stated. – fib4_reader 2 months ago
6Adding for future readers: the certificate should carry the lot number, not just a batch code. – RP_C18 3 months ago
add a comment
Sponsored

Janoshik Analytical - Independent Third-Party Testing

HPLC purity, identity confirmation and quantified content on the vial you actually hold. Reports arrive with the chromatogram attached, not just a number.

Submit a sample
Sponsored — paired listing

GL Biochem (Shanghai) Ltd. - Direct Synthesis

Founded 1998. ISO 9001 and cGMP certified, 1,500+ staff and 200+ patents. The synthesis house behind a great many of the vials that get sent out for testing - batch-specific documentation with every order.

Visit GL Biochem
23

Mechanically, gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

To be exact about it, sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
SI
answeredsample_id17k2724 Jul 2026
5Note that the label instructions differ between agents on precisely this point. – e_dziedzic 3 months ago
add a comment
19

In practice, the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

Worth being precise here: gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
AV
answeredanders_vestby17k2829 Jul 2026
15

Area percentage is not mass percentage, and conflating the two is the most common misreading of a purity figure.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

Compare purity within a single laboratory on the same method, never across laboratories.

edited 20 Jul 2026 by p_mkhize — reworded for clarity after a comment

shareimprove this answerflag
PM
answeredp_mkhize41k13816 Jul 2026
13

On the detail: the short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.

Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
JV
answeredjo_vandeberg19k2717 Jul 2026
4Thank you — the worked example is what makes this usable. – per_haugen 8 months ago
5Related: the same reasoning applies to the counter-ion question. – rota_site 9 months ago
add a comment