How do I reconcile 98.2% from Janoshik with 96.8% from the supplier?

Asked 6 Jun 2025Modified 10 months agoViewed 12k times
3

What I have: 98.2% · Janoshik · 96.8%.

I have a result I cannot explain, and I would rather diagnose it than guess.

I have checked the obvious explanations and eliminated the two easiest ones.

What would you check first, and what would you conclude from each outcome?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

833 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

910 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

771 questions
janoshik
janoshik

Janoshik Analytical, an independent third-party testing laboratory widely used by this community for HPLC purity, identity and quantified content…

139 questions
shareeditfollowflag
TI
askedteodora_ilic17k276 Jun 2025
8Which wavelength was the purity integrated at? Worth adding to the question. – lane_transit 5 months ago
add a comment

5 Answers

Accepted answer first, then by votes
99

Accepted answer

1.4 percentage points, and the direction is the informative half. Janoshik reports 98.2 per cent and the certificate 96.8, so the independent figure is higher. As impurity that is 1.8 per cent against 3.2 — 0.56 times as much unassigned area, which is the unusual direction and worth a second look at the integration. Method differences usually make the independent number the lower one, so an independent result above the certificate points at integration limits, a different wavelength, or a different definition of the main peak rather than at a better vial. Ask both parties for the method section before you decide which figure is wrong, because the answer is often neither.

Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.

Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

Mass shifts and what they usually mean

Δ mass (Da)Most likely causeDistinguishing feature
+1Deamidation (Asn or Gln)New peak, slightly earlier retention
−17Loss of ammoniaOften with deamidation
−18Dehydration / succinimidepH-dependent, reversible
+16Oxidation (Met, Trp)Earlier retention, light-related
−128Missing Gln or LysDeletion sequence from synthesis
0Isomer: racemisation or scramblingSame mass, shifted retention

Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
TG
answered · acceptedtandem_gradient61k24811 Sept 2025
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards
40

Gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

Compare purity within a single laboratory on the same method, never across laboratories.

edited 14 Sept 2025 by a_lindgren — updated for the 2026 guidance change

shareimprove this answerflag
AL
answereda_lindgren58k24831 Aug 2025
5The system-suitability data is the part that tells you whether to believe the rest. – tare_weight 7 months ago
4I would gently push back on the second point — inter-laboratory spread is wider than stated. – tyndall_haze 6 months ago
add a comment
31

Mechanically, the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
BD
answeredb_delacroix43k383 Oct 2025
Confirming from the other direction: I ignored the method section once and paid for it. – ines_brandt 6 months ago
add a comment
25

On the detail: reporting threshold is convention and not chemistry, which is why two certificates with different thresholds disagree by a tenth of a point or more.

Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.

If you only pay for one test, pay for quantified content. Purity is the number everyone quotes and content is the number that changes what you do.

shareimprove this answerflag
TM
answeredtobias_maartens171k35822 Sept 2025
22

Mechanically, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.

Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

In practice: ask for the chromatogram, check the method section, check the lot number against the vial, and set your accept threshold before you see the result rather than after.

shareimprove this answerflag