Quantified content is how many milligrams of peptide are actually in the vial, measured against a calibrated reference standard, and it is the only measurement that directly answers that question.
System suitability for a quantitative method is stricter than for purity because a small systematic error in the standard directly translates into an error in the sample result.
Worth being precise here: if the standard and sample have different absorption coefficients at the detection wavelength, the response factors differ and the inference fails.
The relative standard deviation on replicate quantitations of a homogeneous sample should be below two per cent when the method is under control.
The limitation is that a quantitative method is only as good as the standard it uses, and a cheap standard is a false economy.
Ask for both the purity and the content, and do not accept purity alone.