Mechanically, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
The part that matters: tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
Compare purity within a single laboratory on the same method, never across laboratories.