Accepted answer
The part that matters: the most important parameter is the one that is almost never specified: the gradient slope during the region where your main peak elutes.
Proline-rich sequences are particularly problematic because the isomerisation kinetics are in the same timescale as the separation, leading to split or broadened peaks at low temperature.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Mobile phase pH at the point where you inject must match the mobile phase pH at the start of the gradient, or the sample will not be focused at the column head.
The resolving power of a separation is quantified by the resolution parameter R, defined from the heights and widths of adjacent peaks, and pharmacopoeial methods typically demand R greater than 1.5 for a method to be considered validated.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
If two labs give different numbers, the method difference is the first hypothesis, not lab quality.
3For what it is worth, my own result was within half a per cent of this. – otto_brenner 2 months ago 2Any reason this would differ for a longer peptide? – e_dziedzic 10 days ago add a comment