Accepted answer
It tells you 97.1 per cent of the integrated area came off a C8 column where liraglutide comes off, and the remaining 2.9 per cent did not. That is an area statement at one wavelength, not a mass statement about the vial: 2.9 per cent of area is only 2.9 per cent of mass if every impurity absorbs exactly as strongly as the parent, which none of them do. It also says nothing about how many milligrams are in the glass — water, counter-ion and a short fill are all invisible to it. What a C8 column does add is a constraint on what could be hiding: a column that retains by hydrophobicity separates deletion sequences well and separates isomers of identical hydrophobicity not at all.
The underlying point is that high-performance liquid chromatography is a separation technique, not an identification technique, and it answers the question "how much of the detected signal is a single species" not "which species is it."
Wider-pore phases — 300 Angstrom rather than 100 Angstrom — have faster mass transfer and narrower peaks for peptides above three kilodaltons, which is almost every peptide you will see.
What each test answers
| Test | Answers | Does NOT answer |
|---|
| RP-HPLC, area % | What fraction of detected material is the target | How much target is present |
| Quantified content | Milligrams of peptide per vial | What the impurities are |
| ESI-MS identity | Whether the molecular weight matches | Purity, or isomeric substitution |
| Peptide mapping | Sequence, localised to a fragment | Quantity |
| Karl Fischer | Water content of the solid | Solvent content |
| LAL endotoxin | Pyrogen load in EU/mg | Sterility |
| Sterility test | Growth in defined media over 14 days | Endotoxin, or bioburden count |
To be exact about it, temperature affects both the viscosity of the mobile phase and the dynamics of molecular interactions, and a method developed at 25 degrees and run at 40 degrees will not behave identically.
Peptide impurities from solid-phase synthesis include deletion sequences, truncations from premature cleavage, racemised residues from epimerisation and oxidised variants, each of which may have different chromatographic behaviour.
The practical summary: specify the method, run the same method on every sample you compare, and use orthogonal techniques to confirm the result.