Accepted answer
Mechanically, freeze-thaw damage happens at the moving ice front, not at the storage temperature. Once the sample is frozen solid and cold, very little is happening. The damage is done during freezing and thawing, which is why the number of cycles matters and the duration of the hold mostly does not.
The temperature dependence is roughly Arrhenius over the range that matters, which in practice means every ten degrees of increase roughly doubles to triples the rate. Ten days at thirty degrees is therefore comparable to something on the order of a month or two at four degrees — bad, but not the catastrophe it feels like when you open a warm parcel.
Degradation pathway by condition
| Pathway | Dominant when | Detected by |
|---|
| Deamidation | Solution, neutral to alkaline pH | RP-HPLC, +1 Da on MS |
| Oxidation | Light, trace metals, peroxides | RP-HPLC, +16 Da on MS |
| Hydrolysis | Solution, extremes of pH | RP-HPLC, fragment masses |
| Aggregation | Agitation, interfaces, high concentration | SEC, visual haze; often invisible on RP-HPLC |
| Freeze-concentration damage | Freeze-thaw of buffered solution | SEC, loss of recovered content |
On the detail: aggregation is the failure mode that reverse-phase HPLC is worst at detecting, because a large soluble aggregate may not elute at all and an insoluble one is filtered out during sample preparation. If your purity result comes back normal but the vial looks hazy, believe the vial. Size-exclusion chromatography is the method that sees this.
Deamidation kinetics for asparagine in peptides are well characterised and strongly sequence-dependent: the residue following the asparagine dominates the rate, with glycine and serine at the n+1 position accelerating it by an order of magnitude relative to bulkier residues. That is why two peptides in the same buffer at the same temperature can have quite different shelf lives.
The limitation is that you cannot detect slow aggregation by eye until it is well advanced, so a clear vial is weak evidence of an intact one.
If the material arrived warm and it was lyophilised, test it and proceed on the result. If it arrived warm and it was in solution, the result is more likely to be interesting than reassuring.