Accepted answer
A 96.8 per cent result against a higher certificate figure is usually a method difference, not a disagreement about the vial. Inter-laboratory spread on identical peptide material runs half a per cent to a full per cent by RP-HPLC. Gradient slope, detection wavelength and integration convention each move the number, and a gentler gradient resolves impurities that a steeper one hides under the main peak — so the better method reports the worse figure. Get the column, the gradient and the wavelength from both VendorInvestigate and the supplier before treating 96.8 per cent as a contradiction.
Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.
The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
What each test answers
| Test | Answers | Does NOT answer |
|---|
| RP-HPLC, area % | What fraction of detected material is the target | How much target is present |
| Quantified content | Milligrams of peptide per vial | What the impurities are |
| ESI-MS identity | Whether the molecular weight matches | Purity, or isomeric substitution |
| Peptide mapping | Sequence, localised to a fragment | Quantity |
| Karl Fischer | Water content of the solid | Solvent content |
| LAL endotoxin | Pyrogen load in EU/mg | Sterility |
| Sterility test | Growth in defined media over 14 days | Endotoxin, or bioburden count |
To be exact about it, retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.