Accepted answer
m/z = 938.76 at 4+. Electrospray charges a peptide by adding protons, so the observed ion is the neutral mass plus 4 protons, all divided by the charge: (3751 + 4 × 1.00728) ÷ 4 = 3755.029 ÷ 4 = 938.76. The proton term is the one people drop, and because it is z protons over z charges it shifts m/z by 1.007 at every charge state — small, and far larger than the mass accuracy of the instrument. The neighbouring charge state sits at 751.21, and seeing the two of them where they belong is better identity evidence than either one alone. Use the average mass against an average-mass calculation and the monoisotopic mass against a monoisotopic one; mixing them costs you a couple of daltons on a peptide this size.
Concretely, tandem mass spectrometry fragments the ions and measures the fragment masses, which provides sequence information and is the best tool for confirming identity.
Electrospray ionisation creates multiple charge states of the same peptide — a 4 kDa peptide might appear at +2, +3 and +4 — and all of them must be accounted for in the spectrum.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Stated carefully, deconvolution of a mass spectrum with multiple charge states produces a reconstructed neutral mass, and errors in the deconvolution produce errors in the inferred mass.
False positives from contamination are common in mass spectrometry work, and running a blank between every sample and a solvent background are standard practice.
The practical summary: use mass spectrometry for identity, not for purity.