Accepted answer
0.5 percentage points, and the direction is the informative half. Janoshik reports 97.6 per cent and the certificate 97.1, so the independent figure is higher. As impurity that is 2.4 per cent against 2.9 — 0.83 times as much unassigned area, which is the unusual direction and worth a second look at the integration. Method differences usually make the independent number the lower one, so an independent result above the certificate points at integration limits, a different wavelength, or a different definition of the main peak rather than at a better vial. Ask both parties for the method section before you decide which figure is wrong, because the answer is often neither.
Concretely, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
In practice, tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
2Confirming from the other direction: I ignored the method section once and paid for it. – Dr_Tomas_Kral 4 months ago add a comment