Two people working through the same arithmetic independently should get the same answer, and if they do not, someone has made a unit error.
Dead space quantified: a fixed-needle insulin syringe holds roughly 3 to 5 µL in the hub and needle after the plunger bottoms out. A luer-lock syringe with a detachable needle holds 35 to 100 µL depending on the hub design. At 5 mg/mL that is 15 to 25 µg lost per draw on the insulin syringe and 175 to 500 µg on the luer-lock — which over ten draws is the difference between losing a rounding error and losing half a milligram.
Dead space by syringe type
| Configuration | Dead volume | Loss at 5 mg/mL | Over 20 draws |
|---|
| Fixed-needle insulin syringe | 3–5 µL | 15–25 µg | 0.3–0.5 mg |
| Low-dead-space, detachable | <2 µL | <10 µg | <0.2 mg |
| Standard luer-lock + 30G | 35–60 µL | 175–300 µg | 3.5–6 mg |
| Luer-lock + 21G drawing needle | 70–100 µL | 350–500 µg | 7–10 mg |
Specifically, breaking it down further: if a 10 mg vial has 96.5 per cent content, you have 9.65 mg of peptide. Divide that by 2.00 mL and your concentration is 4.825 mg/mL, not 5.00 mg/mL, which is a 3.5 per cent systematic error in every dose calculation.
The limitation is that technique reduces risk, it does not remove it, and nothing you can do outside a controlled environment makes a non-sterile preparation sterile.
Do the arithmetic twice, ideally with someone else doing it independently.